Please use this identifier to cite or link to this item: https://hdl.handle.net/2440/17310
Citations
Scopus Web of Science® Altmetric
?
?
Type: Journal article
Title: The long-term wear of retrieved McKee-Farrar metal-on-metal total hip prostheses
Author: Howie, D.
McCalden, R.
Nawana, N.
Costi, K.
Pearcy, M.
Subramanian, C.
Citation: Journal of Arthroplasty, 2005; 20(3):350-357
Publisher: Churchill Livingstone Inc Medical Publishers
Issue Date: 2005
ISSN: 0883-5403
1532-8406
Statement of
Responsibility: 
Donald W. Howie, Richard W. McCalden, Namal S. Nawana, Kerry Costi, Mark J. Pearcy, and Chinnia Subramanian
Abstract: Twenty-four cobalt-chrome alloy McKee-Farrar matching acetabular and femoral components were retrieved at revision total hip arthroplasty. The average time in situ was 16 years. Wear and loss of sphericity was very low. Polishing wear (type 1), fine abrasive (type 2), multidirectional dull abrasive (type 3), and unidirectional dull abrasive wear (type 4) of the articulating surfaces were identified. The mean percent area of femoral heads occupied by types 2, 3, and 4 wear was 18%, 5%, and 2%, respectively. There was no association between the type and distribution of wear and the time in situ. Impingement damage was evident on 9 implant pairs. The extent and types of wear described in this paper will be useful when analyzing the patterns of surface damage of newer designs of metal-on-metal articulations.
Keywords: hip arthroplasty
prosthesis
metal-on-metal
wear
electron microscopy
Description: Copyright © 2005 Elsevier Inc.
DOI: 10.1016/j.arth.2004.09.028
Description (link): http://www.elsevier.com/wps/find/journaldescription.cws_home/623141/description#description
Published version: http://dx.doi.org/10.1016/j.arth.2004.09.028
Appears in Collections:Aurora harvest 6
Orthopaedics and Trauma publications

Files in This Item:
There are no files associated with this item.


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.